Adjustment

A4 AGC Gain Adjustment

A4 AGC Gain Adjustment
The purpose of this procedure is to adjust the Gain of AGC on the A4 Receiver Board.

Required Equipment

Nothing

Procedure

Step 1. Run the adjustment program.
Step 2. Choose the A4 Adjustment in the Spot Adjustment.
It is possible to perform this item with the All Adjustment too.
Step 3. Make sure that nothing is connected to the E4991A.
Step 4. Follow the adjustment program instruction to update the correction constant.