Figure 2-7
Segment Table 2 (Power Level: -20dBm, Averaging: 1)
24. Perform OPEN, SHORT, LOAD calibration as described in step 10 to 19.
25. Press
26. Select
27. Press
28. Set OSC[dBm] to -40 and Avg to 1 for all measurement point as Figure 2-8.
Figure 2-8
Segment Table 3 (Power Level: -40dBm, Averaging: 1)
29. Perform OPEN, SHORT, LOAD calibration as described in step 10 to 19.
Chapter 2
to show the Save/Recall softkey menu.
[Save/Recall]
.
Save State
, then E4991A's status is saved to "20.sta" status file.
[-],[2],[0][Enter]
Performance Test and Function Test
Performance Test
37