Troubleshooting
Booting Process Troubleshooting (for serial prefix MY433 and later)
Figure 4-17
Measurement view
Step 4. Power On Self Test
The power on self test is executed once automatically before the measurement starts. For
details, refer to "Troubleshooting Using Internal Test" on page 172. In addition, a message
of the "Internal Test In Progress..." is displayed during the power on self test in progress. If
an hang-up occurs during the power on self test, problem seems in analog block such as A1
Reference OSC, A3 Source, A4 Receiver or A2 DC Bias(Opt.001).
Chapter 4
171