Performance Test and Function Test
Performance Test
63. Record the |Z| and  readings to the calculation sheet.
NOTE
Ignore the reading at 1.3 GHz and 2.8 GHz.
64. Calculate the test result according to the calculation sheet, the record it to the
10 cm airline with SHORT Termination
65. Record the 16190B's 10 cm airline with SHORT termination calibration values on the
66. Connect the 16190B's 10 cm airline and short termination to the 7mm connector, using
Figure 2-10
10 cm Airline with SHORT Measurement Test Setup
67. Press
68. Select
69. Select the file "0.sta".
40
performance test record.
calculation sheet.
the following procedure. (see Figure 2-10)
a. Remove the open termination from the airline.
b. Gently inserts the airline center conductor into the short termination center
conductor. Mate the outer conductors. Then torque the connection to 136 N-cm
using the Torque Wrench(p/n:8710-1760). A 1/2 inch open end wrench may be
necessary to hold the airline stationary.
to show the Save/Recall menu.