Post-Repair Procedures
Post-Repair Procedures
Table 8-1
Post-Repair Procedures
Replaced
Required Adjustments
Assembly or Part
Correction Constants (CC)
A22 Front Panel
Not Required
Keyboard
A24 GPIB Board
Not Required
Not Required
A25 LAN Interface
Board
Not Required
A26 Display Board
A27 Mass Storage
All Adjustments.
Disk Drive
System Backup
Not Required
A28 FDD
A50 Power Supply
Not Required
Assembly
Not Required
A51 LCD
A52 Inverter Board
Not Required
CPU 400MHz
Not Required
DIMM 64Mbyte
Not Required
EEPROM on the
Write ID (Serial No. and Option No.)
A21
EEPROM Initialize
All Adjustments.
System Backup
310
Verification
"To Execute the External Test" on page 189
"To check the Front Panel" on page 179
"To check the GPIB" on page 188
"To check the LAN (for serial prefix JP1KH,
JP2KH, and MY432)" on page 181
Inspect the Booting Process
"To Execute the External Test" on page 189
"To check the VGA Display Output" on
page 180
Inspect the Booting Process
Perform the all Performance Tests
"To check the FDD" on page 180
Inspect the Booting Process
Inspect the Booting Process
"To Execute the External Test" on page 189
"To check the LCD" on page 179
Inspect the Booting Process
"To Execute the External Test" on page 189
Inspect the Booting Process
Inspect the Booting Process
Inspect the Booting Process
Perform the all Performance Tests