Dc Bias Level Adjustment; Dc Bias Level Adjustment Setup - Agilent Technologies 4284A Service Manual

Precision lcr meter
Hide thumbs Also See for 4284A:
Table of Contents

Advertisement

2-1 0. DC BIAS LEVEL ADJUSTMENT (OPTION 001 ONLY)
This adjustment updates the E E PROM Calibration Data to set the DC bias voltage. The
adjustment program is necessary to perform this adjustment.
EQU IPMENT:
Computer
M ultimeter
H P-I B Cable
Interface Box
B N C(m) to BN C(m) Cable ( 30 cm )
BNC(f) to Dual Banana Plug Adapter
PROCEDURE:
1 . Display the adjustment program's main menu , referring to paragraph 2-7-
2. I N ITIAL OPERATI NG PROCE D U R E .
2 . Press the 'Adj. 1 ' softkey t o select the DC Bias Level Adjustment.
When the computer cannot access the H P 3458A, the H P- I B address
registration display appears . Enter the HP 3458A's H P- IB add ress
using the numeric keys and the 'Enter' softkey.
3. Perform the adjustment acco rding to the program ' s instruction.
shows the adjustment setup.
MU LTIMETER
oo oooc o
OO����O
O DOClOOO
00 0 0
CJClDDDD
N OTE
BNC(f)-dual banana
plug Adapter
000
BNC(m)-BNC(m)
Cable (30 cm)
Figure 2- 1 0 . DC Bias Level Adjustment setup
2-1 6
H P 9826A
HP 3458A
H P 1 0833A
HP PN 04284-65007
HP PN 8 1 20-1 838
HP PN 1 25 1 -2277
Figure 2-1 0
o
o
HP
o
428 4A
ODD
o
0 0 000
0
o 0
=
INTERFACE BOX
BIAS
TEST SIGNAL
DC
0
0