Step 4. Click OK, then wait till the measurement completion.
Step 5. Disconnect the power sensor and the adapter from the 7mm connector.
Step 6. Connect the instrument as shown in Figure 3-13.
Figure 3-13
A3 Output Level Adjustment Setup (2/2)
Step 7. Follow the adjustment program instruction to update the correction constant.
Chapter 3
Adjustment
A3 Output Level Adjustment
137