Agilent Technologies E4991A Service Manual page 143

Rf impedance/material analyzer
Hide thumbs Also See for E4991A:
Table of Contents

Advertisement

Figure 3-16
A5 Reference Oven Adjustment Setup
Step 4. Follow the adjustment program instruction to update the correction constant.
Step 5. Adjust "Reference Oven Adjustment screw" as shown in Figure 3-17 so that the counter
reading is 10MHz
strap handle.
Figure 3-17
Reference Oven Adjustment Screw
Chapter 3
A5 Reference Oven Adjustment (Opt 1D5 only)
1Hz, if the counter reading is out of 10MHz
Adjustment
1Hz. Then remove side
143