Figure 2-4
Power level Accuracy Test Setup 2
12. Set the E4991A as follows;
Setting
Frequency: 1GHz
Power: 1dBm
13. Input the E4991A's frequency to the power meter.
14. Press
15. Subtract theE4991A's reading from the power meter reading, then record it to the
performance test record.
16. Set the OSC level and Frequency settings as shown in Table 2-1 to test the analyzer and
repeat step 12 to 15 for other points.
Chapter 2
Operation
-
[start/stop]
Center
[Source] - Osc Unit- Power
[Source] - Osc Level- [1]-[Enter]
to make a measurement.
[Trigger]
Performance Test and Function Test
Performance Test
-
[1]-[G/n]
33