A4 Local Level Adjustment

The purpose of this procedure is to adjust the local level in the A4 Receiver Board.

Required Equipment

Nothing

Procedure

Step 1. Run the adjustment program.
Step 2. Choose the A4 Adjustment in the Spot Adjustment.
It is possible to perform this item with the All Adjustment too.
Step 3. Make sure that nothing is connected to the E4991A.
Step 4. Follow the adjustment program instruction to update the correction constant.
Chapter 3
Adjustment
A4 Local Level Adjustment
139