Adjustment
Adjustment Program

How to perform the Adjustment Program

The following is the simple procedure to perform the Adjustment Program. For details,
refer to the "Operation Manual for E4991A Adjustment Program".
Step 1. If the following equipment is used for the E4991A Adjustment, you need to use Agilent
Edit Cal(Calibration Data Editor).
If you don't use the following equipment, go to step 3.
Install the latest version: Rev.A.02.30(as of November 2001)of Agilent Edit Cal before
running the Adjustment Program, if it is not installed in your PC. For details, refer to
"Operation Manual for Edit Cal". The program and operation manual for the Edit Cal are
downloadable from CTU-Kobe web site.
Step 2. Enter the calibration data of the Equipment(8482A) in the Edit Cal.
Step 3. Start the Agilent VEE.
Step 4. Load the adjustment program file into the Agilent VEE as follows.
1. Pull down the File menu from the
2. Select the file C:\adje4991a\adje4991a.vee and click
Step 5. You may be asked to add drivers for the equipment during the program loading. Then, click
OK
which are not used for the adjustment. (See Figure 3-4).
Figure 3-4
Direct I/O configuration
Step 6. Click
128
8482A Power Sensor
and the GPIB address for each equipment. Enter 0 as the address for the equipment
on the Agilent VEE Screen as shown in Figure 3-5.
START
VEE window and select
Agilent
.
Open
.
Open
Chapter 3