4349B
6. Press
to perform the OPEN correction.
7. Set up the equipment as shown in Figure 9-5. If the DC Voltage Calibrator listed in
Table 9-2 is available, setup the equipment as shown in Figure 9-6.
Figure 9-5. Current Measurement Accuracy Test Setup Using DC Power Supply
Figure 9-6. Current Measurement Accuracy Test Setup Using DC Voltage Calibrator
Maintenance
9-7