a. Fully retract the threads on the 7mm connector. Then insert the marked side tip of
b. Gently cover the airline center conductor with the airline outer conductor, with the
c. Gently inserts the airline center conductor into the open termination center
Figure 2-9
10 cm Airline with OPEN Measurement Test Setup
57. Press
58. Record the |Z| and  readings to the calculation sheet.
NOTE
Ignore the reading at 1.3 GHz and 2.8 GHz.
59. Press
60. Select
61. Select the file "-20.sta".
62. Press
Chapter 2
the airline center conductor.
logo side down. (To prevent damage, don't let the conductor scrape the edge of the
outer conductor.) Mate the outer conductors. then torque the connection to 136
N-cm. (A 1/2 inch open end wrench may be necessary to hold the airline
stationary.)
conductor. Mate the outer conductors. Then torque the connection to 136 N-cm
using the Torque Wrench(p/n:8710-1760). A 1/2 inch open end wrench may be
necessary to hold the airline stationary.
to make a measurement.
[Trigger]
to show the Save/Recall menu.
[Save/Recall]
.
Recall State
to make a measurement.
[Trigger]
Performance Test and Function Test
Performance Test
39