Performance Test Record (Option 007)

OSC Level Accuracy Test (Option 007)
Power Level
Frequency
1 dBm
1 GHz
0 dBm
3 GHz
7.9 dBm
3 GHz
10 dBm
300 MHz
15 dBm
1 MHz
20 dBm
50 MHz
25 dBm
1 GHz
30 dBm
10 MHz
35 dBm
2 GHz
40 dBm
100 MHz
Chapter 2
Test Limit
Test Result
-4.00/+2.00 dB
-6.00/+3.00 dB
-6.00/+3.00 dB
-4.00/+2.00 dB
-4.00/+2.00 dB
-4.00/+2.00 dB
-4.00/+2.00 dB
-4.00/+2.00 dB
-6.00/+3.00 dB
-4.00/+2.00 dB
Performance Test and Function Test
Test Record
Measurement
Uncertainty
0.22 dB
dB
0.35 dB
dB
0.35 dB
dB
0.26 dB
dB
0.22 dB
dB
0.24 dB
dB
0.24 dB
dB
0.22 dB
dB
0.24 dB
dB
0.27 dB
dB
TUR
9.44
8.72
8.62
7.88
9.40
8.57
8.57
9.38
12.69
7.63