3. Test Port Output Power Level Accuracy
Perform this test to confirm the accuracy of the 8753ES test port output power. A power
meter is used to determine the output level from Port 1. This measurement is compared to
the level set by the network analyzer. The difference must be within the specified
tolerance.
Analyzer warm-up time: 30 minutes
Specifications
Frequency Range
300 kHz to 3 GHz
b
3 GHz to 6 GHz
a. At 0 dBm and 25° C ±5° C
b. Only for analyzers with Option 006
Required Equipment
Description
Power Meter
Power Sensor
Adapter: APC-7 to Type-N (f)
Additional Equipment for Option 006 Analyzers
Power Sensor
Additional Equipment for 75Ω Analyzers
Power Sensor
Procedure
1. Zero and calibrate the power meter. For more information of how to perform this task,
refer to the power meter operating manual.
2. Connect the equipment as shown in
Chapter 2
Agilent 8753ES System Verification and Performance Tests
Test Port Output Power
a
Accuracy
±1.0 dB
±1.0 dB
HP/Agilent Model Number
436A/437B/438A or E4418B/4419B
8482A
11524A
8481A
8483A Option H03
Figure
2-10.
System Verification and Performance Tests
2-19