Standard:
Averaging:
Frequency
Power
Level
1 MHz
-3 dBm
10 MHz
-3 dBm
100 MHz
-3 dBm
200 MHz
-3 dBm
300 MHz
-3 dBm
500 MHz
-3 dBm
600 MHz
-3 dBm
800 MHz
-3 dBm
1.0 GHz
-3 dBm
1.3 GHz
-3 dBm
1.6 GHz
-3 dBm
1.8 GHz
-3 dBm
2.0 GHz
-3 dBm
2.2 GHz
-3 dBm
Chapter 2
50
8
Measurement
Test Limit
Parameter
0.35
|Z|
7.0 mrad
0.36
|Z|
7.2 mrad
0.43
|Z|
8.5 mrad
0.58
|Z|
11.5 mrad
0.65
|Z|
13.0 mrad
0.80
|Z|
16.0 mrad
1.08
|Z|
21.5 mrad
1.23
|Z|
24.5 mrad
1.38
|Z|
27.5 mrad
2.25
|Z|
45.0 mrad
2.48
|Z|
49.5 mrad
2.63
|Z|
52.5 mrad
4.02
|Z|
80.5 mrad
4.17
|Z|
83.5 mrad
Performance Test and Function Test
Test Result
Measurement
Uncertainty
0.09
1.7 mrad
mrad
0.09
1.7 mrad
mrad
0.09
1.7 mrad
mrad
0.11
2.2 mrad
mrad
0.14
2.7 mrad
mrad
0.19
3.7 mrad
mrad
0.23
4.6 mrad
mrad
0.23
4.6 mrad
mrad
0.23
4.6 mrad
mrad
0.24
4.7 mrad
mrad
0.24
4.8 mrad
mrad
0.24
4.6 mrad
mrad
0.51
10.1 mrad
mrad
0.51
10.2 mrad
mrad
Test Record
TUR
4.17
8.87
4.26
8.87
5.09
8.87
5.30
8.87
4.92
8.87
4.44
8.87
4.70
8.87
5.37
8.87
5.99
8.87
9.60
8.87
10.43
8.87
11.32
8.87
8.00
8.87
8.20
8.87
105