System Verification and Performance Tests
Performance Tests
Level Accuracy Performance Test
This test checks the maximum variation in output power across the frequency range of the
analyzer.
Table 2-7 Required Equipment
Equipment Description
HP/Agilent Power Meter
HP/Agilent Power Sensor
Adapter 3.5-mm (f) to 3.5-mm (f)
Adapter 2.4-mm (f) to 2.4-mm (f)
Analyzer warmup time: 30 minutes.
1. Zero and calibrate the power meter.
2. Set the calibration factor on the power meter to the average value of the power sensor
between 50 MHz and 20 GHz (40 GHz for 8722ET/ES).
For example, if the power sensor calibration factor is 100% at 50 MHz and 92% at
20 GHz (40 GHz for 8722ET/ES), set the calibration factor to 96%.
3. Connect the equipment as shown in
Figure 2-18 Power Test Setup
2-36
8719ET/ES
8720ET/ES
436A, 437B,438A E4418A (EPM-441A) or
E4419A (EPM-442A)
8485A
P/N 85052-60012
(part of 85052B)
N/A
Figure
2-18.
8719ET/20ET/22ET
8719ES/20ES/22ES
RF Network Analyzers
8722ET/ES
8487A
N/A
P/N 85056-60006
(part of 85056A)
Service Guide