Agilent Technologies E4440AU Installation Note

Agilent psa spectrum analyzer option 071 digital demod hardware retrofit kit for serial prefix us4142 and my4100

Advertisement

Quick Links

Installation Note
Agilent PSA Spectrum Analyzer
E4440AU Option 071 Digital Demod
Hardware Retrofit Kit
for Serial Prefix < US4142 and < MY4100
Part Number E4440-90596 Supersedes: E4440-90577
Printed in USA June 2005

Advertisement

loading

Summary of Contents for Agilent Technologies E4440AU

  • Page 1 Installation Note Agilent PSA Spectrum Analyzer E4440AU Option 071 Digital Demod Hardware Retrofit Kit for Serial Prefix < US4142 and < MY4100 Part Number E4440-90596 Supersedes: E4440-90577 Printed in USA June 2005...
  • Page 2 The information contained in this document is subject to change without notice. Agilent Technologies makes no warranty of any kind with regard to this material, including but not limited to, the implied warranties of merchantability and fitness for a particular purpose.
  • Page 3 Digital Demodulation Hardware Retrofit Kit Digital Demodulation Hardware Retrofit Kit Product Affected: PSA E4440A Serial Numbers: US00000000/US4042xxxx MY00000000/MY4042xxxx To Be Performed By: (X) Agilent Service Center (X) Personnel Qualified by Agilent ( ) Customer Estimated Installation Time: 1 Hour Estimated Adjustment and Verification Time for Hardware Kit Following Installation: 6.5 Hours (see note 1) Additional Recommended Task...
  • Page 4 Digital Demodulation Hardware Retrofit Kit Contents Item Description Agilent Part Quantity Number Option Upgrade Entitlement Certificate Basic Mode Guide PSA Documentation CD Rom A27 Electronic Attenuator E4440-60432 W51, Cable, semi-rigid, E-Attenuator to RYTHM E4440-20110 W52, Cable, semi-rigid, E-Attenuator to 3 GHz LPF E4440-20109 W53, Cable, ribbon, E-Attenuator to Front End Driver E4440-60078...
  • Page 5: Installation Procedure

    Installation Procedure Installation Procedure 1. Perform the following procedures in order: a. Remove the instrument outer case. b. Remove the instrument top brace. c. Install the Option 071 retrofit kit. d. Install the instrument top brace and outer case. e. Install the option designator key. f.
  • Page 6 Installation Procedure Figure 1 Instrument Outer Case Removal Installation Note E4440-90596...
  • Page 7 Installation Procedure Remove the Top Brace 1. Refer to Figure 2. Use the T-10 driver to remove the top screws (3) (one screw is under the security label), and the side screws (2) attaching the top brace (1) to the deck.
  • Page 8 Installation Procedure Attach the Attenuator to the Bracket 1. Refer to Figure 3. Install the electronic attenuator onto the attenuator bracket using four of the screws contained in the kit; torque to 101 Ncm (9 inch pounds). Figure 3 Attach the Attenuator to the Bracket Installation Note E4440-90596...
  • Page 9 Installation Procedure Install the Attenuator Figure 4 shows the RF section and parts locations. Locate the RF section on the right side of the instrument. Figure 4 Remove Cables In the following step 2, use a 7/16-inch wrench on one of the wrench CAUTION flats at the ends of the filter to prevent the filter body from rotating when removing the cable.
  • Page 10 Installation Procedure Figure 5 Install the Attenuator 5. Select the E4440-20110 semi-rigid cable from the kit. This cable is reference designator W51, shown in Figure 5. Connect the cable between the RYTHM and the electronic attenuator. Torque to 112 Ncm (10 in-lb). 6.
  • Page 11 Installation Procedure Replace the Top Brace and Outer Case 1. Refer to Figure 2. Carefully position the top brace on the deck. The alignment pin at the center of the web/fan assembly must mate with the alignment hole on the top brace.
  • Page 12 Installation Procedure Obtaining a License Key and Activating the Option The entitlement certificate supplied in this kit allows you to obtain a license key from our Agilent website so you can enable this upgrade option. Once you have retrieved the license key, you can begin the process of activating the option.
  • Page 13 Installation Procedure Initialize Instrument Memory NOTE This procedure requires the use of software included in the PSA Series Calibration Application Software. 1. Load the “PSA Series Calibration Application Software” on your PC. 2. Configure the unit under test by selecting PSA model number E4440A, and press the Configure UUT button.
  • Page 14 Frequency Response Above 3 GHz Noise Sidebands < 50 kHz Offsets Noise Sidebands > 50 kHz Offsets For assistance, contact your nearest Agilent Technologies Sales and Service Office. To find your local Agilent office access the following URL: http://www.agilent.com/find/assist Installation Note E4440-90596...