Required Adjustment After Replacing Assembly - Agilent Technologies E4991A Service Manual

Rf impedance/material analyzer
Hide thumbs Also See for E4991A:
Table of Contents

Advertisement

Adjustment

Required Adjustment after Replacing Assembly

Required Adjustment after Replacing Assembly
If you replace the following assembly, the adjustment items as shown in Table 3-2 are
required. Then perform the corresponding adjustment using the Spot Adjustment in the
Adjustment Program.
Table 3-2
Required Adjustment Item after Replacing Assembly
Replaced Assembly
A1 REFERENCE OSCILLATOR
A3 SOURCE
A4 RECEIVER
A6 STANDARD TEST HEAD
A2 DC Bias(Opt. 001)
A5 10MHz OVEN(Opt.1D5)
*1.If your instrument has an option 001(A2 DC Bias), also perform A2 Adjustment.
124
Required Adjustment
(Spot Adjustment in Adjustment Program)
*1
*1
*1
Chapter 3