Replacement Procedure

Replacing an Assembly

Replacing an Assembly
The following steps show the sequence to replace an assembly in a E4991A RF
Impedance/Material Analyzer.
1. Identify the faulty group. Refer to Chapter 4, "Troubleshooting."
2. Order a replacement assembly. Refer to Chapter 6, "Replaceable Parts."
3. Replace the faulty assembly and determine what adjustments are necessary. Refer to
4. Perform the necessary adjustments. Refer to Chapter 3, "Adjustment."
5. Perform the necessary performance tests. Refer to Chapter 2, "Performance Test and
WARNING
These servicing instructions are for use by qualified personnel only. To avoid
electrical shock, do not perform any servicing unless you are qualified to do so.
WARNING
The opening of covers or removal of parts is likely to expose dangerous voltages.
Disconnect the instrument from its power supply.
CAUTION
Many of the assemblies in this instrument are very susceptible to damage from
ESD(electrostatic discharge). Perform the following procedures only at a static-safe
workstation and wear a grounding strap.
284
Chapter 7, "Replacement Procedure." and Chapter 8, "Post-Repair Procedures."
Function Test."
Chapter 7