Connecting The Injection Valve Waste Line - Thermo Scientific Dionex ICS-5000+ Installation Instructions Manual

Ion chromatography system
Table of Contents

Advertisement

3.7.2

Connecting the Injection Valve Waste Line

For a Dionex AS-DV
1. Cut a length of 0.25 mm (0.010 in) ID black PEEK tubing (analytical,
P/N 042690; capillary, P/N 082647) long enough to reach from the
injection valve to the waste container.
2. Connect one end to port 6 on the injection valve.
NOTE
3. Route the other end to waste.
For a Thermo Scientific Dionex ICS Series AS-AP Autosampler
1. For a capillary system:
Capillary
IC
a. Locate the following items in the Dionex AS-AP Ship Kit
(P/N 074929):
b. Cut a 61 cm (24 in) length of black tubing. If this is not long enough
to reach from the injection valve to the Dionex AS-AP waste port,
add an additional length of green PEEK tubing.
2. For an analytical system:
a. Locate the system tubing package supplied with the injection valve.
One tubing package with precut and labeled tubing (either microbore
or standard bore) is provided for each valve installed in the DC.
b. In the system tubing assembly bag, locate the 0.75 mm (0.030 in) ID
green PEEK tubing assembly, labeled
NOTE
Doc. 065447-02 12/14
In a capillary system, port 6 is labeled
In an analytical system, port 6 is labeled
0.75 mm (0.030 in) ID green PEEK tubing (P/N 044777)
0.25 mm (0.010 in) ID black PEEK tubing (P/N 042690, for
analytical; P/N 082647, for capillary)
10-32 fitting bolt (P/N 074449) and 10-32 double-cone ferrule
(P/N 074373)
To improve sample loading precision in analytical sys-
tems, add a 51 cm (20 in) piece of black PEEK tubing
to the green waste line.
3 • System Setup
.
SAMPLE OUT
.
W
TO INJ VALVE-W.
43

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents