Chapter 5. Electrical Requirements

5.1 Power Requirements

5.1.1 Electrical Requirements

Innova System requires a specific power line, with three phase (see Illustration in
on page
140).
Table 5-2 Core system
Nominal voltage
380 V ± 10 %
400 V ± 10 %
50 Hz or 60 Hz (± 3 Hz)
415 V ± 10 %
480 V ± 10 %
The Hospital circuit breaker should fit the current protection of the Innova system:
• 150A/ 480V, 3 Phases for UL
• 80A/ 380V/400V/415V, 3 Phases for CE
Table 5-3 Options
Option
Nominal voltage
LDM
100-120 V / 220-240 V 50 Hz or 60 Hz (± 3 Hz)
AW
100-127 V / 200-240 V
S5I GE
100-120 V / 230 V
NOTE
PDB maximum rating is equal to 211 kVA.
Large Display Option requires an additional power line for Large Display Cabinet (see illustration in
5.1.2 Power Distribution on page
This power line should fit the current protection requirement of Large Display Cabinet:
• 30A /120V / AWG10 (UL option)
• 30A /120V / AWG10 (CE option)
5750182-1EN 3
Frequency
Nominal (at 125 kV,
100 ms, 640 mA)
60 Hz (± 3 Hz)
Frequency
50 Hz or 60 Hz
50 Hz or 60 Hz
140).
TM
Innova
IGS 6
Power consumption
Peak
60 kVA
150 kVA
Nominal Power consumption Type of power input
3 kVA
11 A / 5.5 A
400 VA
Electrical Requirements
5.1.2 Power Distribution
Type of power input
Without Fluoro
With Fluoro
UPS
~
3
Single phase
Single phase
Single phase
UPS
~
3N
137