Peripheral/Option Checks

If any peripherals or options are not part of the system configuration, the check can be omitted.
Refer to
Table 10-6
Step
1
2
3
4
5
6
7
8
9
10-5-3
Input Power
10-5-3-1
Mains Cable Inspection
Table 10-7
Step
1
2
3
4
10-5-4
Cleaning
10-5-4-1
General Cleaning
Frequent and diligent cleaning of the Voluson® 730Pro ultrasound system reduces the risk of spreading
infection from person to person, and also helps to maintain a clean work environment.
Table 10-8
Step
1
2
10-8
DRAFT (A
Table 3-7, "Approved Peripherals," on page 3-43
Approved Peripheral/Hardware Option Functional Checks
Item
VCR
Verify record/playback capabilities of the VCR. Clean heads and covers if necessary.
B/W Printer
Verify hardcopy output of the B/W video page printer. Clean heads and covers if necessary.
Color Printer
Verify hardcopy output of the Color video page printer. Clean heads and covers if necessary.
Line Printer
Verify hardcopy output of the Line printer. Clean heads and covers if necessary.
DICOM
Verify that DICOM is functioning properly. Send an image to a DICOM device.
Footswitch
Verify that the footswitch is functioning as programed. Clean as necessary.
DVD and/or
Verify that the DVD-drive and/or optional MO-drive reads/writes properly
MO-Drive
(export/recall images in Sonoview).
Modem
Verify modem remote connection.
ECG
Verify basic operation with customer.
Mains Cable Inspection
Item
Unplug Cord
Disconnect the mains cable from the wall and system.
Inspect
Inspect it and its connectors for damage of any kind.
Verify that the LINE, NEUTRAL and GROUND wires are properly attached to the terminals, and that
Verify
no strands may cause a short circuit.
Verify
Inlet connector retainer is functional.
General Cleaning
Item
Use a fluid detergent in warm water on a soft, damp cloth to carefully wipe the entire system.
Console
Be careful not to get the cloth too wet so that moisture does not enter the console.
Caution: DO NOT allow any liquid to drip or seep into the system.
Clean Top surface, Monitor and Touch Panel with a fluid detergent in warm water on a soft, damp cloth.
Monitor
Caution: DO NOT spray any liquid directly onto the Covers, Monitor, keyboard, etc.
Section 10-5 - System Maintenance
28, 2008)
PRIL
for a list of approved peripherals.
Description
Description
Description
V
® 730P
/ P
OLUSON
RO
RO
S
ERVICE
V (BT03)
M
ANUAL

Advertisement

Table of Contents
loading

This manual is also suitable for:

Voluson 730prov

Table of Contents