GE HealthCare CONFIDENTIAL
Direction 5936427, Rev. 1
LOGIQ Totus Basic Service Manual
4-3-8

B-Mode functional checks

B-Mode is the Ultrasound system's default mode.
For information on the Ultrasound system's different modes as well as expected results, Chapter 5 of
the LOGIQ Totus Basic User Manual (latest revision), or the appropriate LOGIQ Totus Release Notes.
1-4-1 "Label Location" on page
1-13.
Figure 4-12 B-Mode Screen Example
Depending on probe availability, choose the preset application shown below. The LOGIQ Totus probes
are ONLY approved for use in the applications listed in Chapter 11 of the LOGIQ Totus Basic User
Manual (latest revision), or the appropriate LOGIQ Totus Release Notes.
IQC preset for Service is available
Image Quality Check (IQC) is intended to facilitate image quality checks during quality assurance
evaluations. Quality assurance tests are used to determine whether a scanner is providing the same
level of performance year after year.
By using the same settings year after year, this ensures that the data collection is consistent
independently of who performs the test.
This preset only includes fundamental settings for B-Mode.
Processing modes like SRI, Harmonics, etc., are turned off.
To do an Image Quality Check (IQC):
1.) Navigate to Utility > Imaging Preset Manager.
2.) Select the probe.
3.) Under Imaging Preset Selections, select Category.
4.) Expand the plus sign in front of IQC for Service.
5.) Select IQC and then the right arrow to assign it to a touch panel key.
6.) Map the IQC to the location you want it to appear on the touch panel.
7.) Select Probe > IQC.
Chapter 4 Functional Checks
4 - 19