Start Again
System Zero
By Cert.
Advanced
Mantracourt Electronics Limited
Click this button when the input has been applied and the reading has been
allowed to settle. This will acquire the reading and allow you to move to the
next points. You will be able to click the button again to re-acquire.
Click here to restart the calibration.
Once calibrated you may want to remove a fixed system value. In the case of a
strain gauge input this may be the weight of a sling, shackle, load bed etc.
Apply the required input and click here to set the system zero. The current input
will be removed from subsequent readings so that the reading will be zero.
To edit this value manually click the Advanced button.
System Zero is stored in non-volatile memory in the transmitter module.
You can click the By Cert button to calibrate against a sensor calibration sheet.
You just need to enter the input values and associated engineering unit
required output value of at least 2 points.
This will take you to a different screen.
Clicking the advanced button will allow you to edit the gains and offsets for
each available calibration point. This will take you to a different screen.
98
T24 Telemetry User Manual