Zero Indication Band
Allow System Zero
Perform System Zero
Mantracourt Electronics Limited
Using this setting you can mask tiny changes in input after you press the Tare
button.
Entering zero will disable this function.
Entering a non-zero value will provide a band within which the display will always
read zero.
Once the reading exceeds this value the real weight will be displayed as no taring
is taking place.
Example: You are adding boxes to a platform and you press tare between adding
each one so you can see the weight of each box.
Without this setting activated each time you tare the display will be around zero
but not exactly zero (By setting the display resolution you may hide this
difference) by setting a small value here such as 0.2kg the display will show a
stable zero while actual weight is fluctuating less than ± 0.2kg.
Entering a non-zero value here will enable system zero to be performed by
holding down the Tare key for a number of seconds.
The value entered here represents the number of seconds the Tare key needs to
be held.
This section allows the user to apply or remove a system zero.
This will require that the transmitter modules are configured and attached to the
handheld and the entire system is ready for zeroing.