Indicators

G
NET
SIG LOW
BATT LOW
REMOTE ERROR
REMOTE BATT LOW

Errors

Displayed on handheld LCD.
Error 1
Error 2
Overload
Mantracourt Electronics Limited
The display is showing Gross load.
The display is showing Net load.
The radio signal from the transmitter module is low. The module is still
functioning but the limit of the range may be near. Communications may start
to deteriorate when this indicator is visible. Until ------ is displayed the
communications is still OK and the display can be relied on for accuracy.
Even with a degraded signal the display value will always be
correct.
The batteries in the handheld are low and need to be replaced.
The transmitter module has an error that the handheld does not recognise.
The battery or supply to the transmitter module is low.
The transmitter module has a strain gauge input and is in shunt calibration
mode. An external module has placed the transmitter module in Shunt
Calibration mode so rather than display a misleading reading this error is
displayed instead.
Input integrity error. The transmitter module has found a problem with the
input. There may be open or short circuits. Rather than display a misleading
reading this error is displayed instead.
The overload limit set by the user has been exceeded.