Zero Settings - Mantracourt T24 User Manual

Hide thumbs Also See for T24:
Table of Contents

Advertisement

Zero Settings

Here you can adjust settings that affect the display of zero.
Items you can change:
Power On Auto Zero
Mantracourt Electronics Limited
Here you can determine whether the T24-HA performs automatic zero when it is
powered on.
Enter zero to disable this function.
If you enter a non-zero value then when the handheld is first turned on it checks
the value read from the transmitter module. If this falls within ± of this value then
the display will be altered so this reads zero.
Example: A strain gauge transmitter module (T24-SA) is calibrated in kg and
measures the weight of boxes on a platform. The weight of the platform itself has
been removed using system zero on the transmitter module.
Sometimes there is debris on the platform which you do not want to see when
viewing the weight of boxes that will be placed on the platform later.
The minimum weight of a box is 5 kg so you could set the Power On Auto Zero to
2 kg.
When you turn on the handheld, if the weight on the platform is between -2 and
+2kg then the handheld will tare this weight off and so read zero.