Configuration

The T24 Toolkit provides a means of simple configuration and calibration of the transmitter module along with
useful tools to aid integration.
Launch the T24 Toolkit software application and pair to this module to enable the connection to the Toolkit to
allow configuration to take place.

Data Rates and Quality

This page allows you to select the rate at which data is transmitted from the transmitter module and the quality.
By selecting low power mode and entering some other information the toolkit will also give guides on achievable
battery life.
Note that the battery life calculator is assuming the best case scenario which is at 20°C and that the battery has a
suitable low internal resistance or that a suitable capacitor is fitted across the battery. See battery details in the
Installation section.
Items you can change:
Transmit Interval
Sample Time
Mantracourt Electronics Limited
See Common Toolkit Pages - Home
Enter the transmission rate in milliseconds. The default is 300 giving
approximately 3 per second which is ideally suited to reading on a handheld.
You may want to slow this down to achieve longer battery life.
This is the length of time in milliseconds that the input is sampled before the
value is transmitted. This can vary between 5 milliseconds and close to the
Transmit Interval. A shorter sample time means that the module is awake for
less time so battery life is increased but at the expense of a reading with less
noise free resolution. You can vary this to see the effect on battery life and noise
free resolution.