Radio Range; Battery Types - Mantracourt T24 User Manual

Hide thumbs Also See for T24:
Table of Contents

Advertisement

Power Supply
Standby / Low Power Mode
Normal Mode on constantly
Reverse Polarity Protection
Internal
Battery Supply Voltage
Current
External
Power Supply voltage
Power Supply ripple
Current
2.
Power supply must be capable of supplying 300 mA for 250 µs
Battery Life in Low Power Mode
Generating Results every 2 seconds
Pair D cells
Pair D cells

Radio Range

To determine radio range please refer to

Battery Types

Battery Type
Lithium Iron Disulphide
Li-FeS
2
Alkaline
Zn-MnO
2
Nickel Metal Hydride
NiMh
Nickel Cadmium
NiCad
Mantracourt Electronics Limited
Min
2.1
Usage
Constantly on
12 sessions per day of 5 minutes
Appendix B – Antenna Range
Notes
These can be found at 1.5 volts in AA size and can therefore be a direct replacement
for Alkaline cells. The low internal resistance and high capacity make these batteries
an ideal choice. The shelf life is around 20 years.
Recommended for AA battery powered modules: Energizer Ultimate Lithium L91
Pairs of alkaline 1.5 V cells are the most common. Use D cells for maximum life and
AA cells where space is restricted. Typical capacity is 2Ah.
Example: Varta 4014 (D), Varta 4006 (AA)
Most cells are 1.2 V so two in series gives 2.4 V. These can match alkaline batteries in
capacity but as the charged voltage is lower they do not match the usable capacity.
These batteries self discharge at a faster rate than alkalines. If charging these cells in
circuit precautions must be taken to ensure that the maximum voltage on the
transmitter module is not exceeded.
Example: GP 270AAHC (AA)
Most cells are 1.2 V so two in series gives 2.4 V. Three in series can be used to give
3.6 volts. These do not have the usable capacity of an alkaline battery. These are
generally only useful if they are to be charged on a regular basis. If charging these
cells in circuit precautions must be taken to ensure that the maximum voltage on the
transmitter module is not exceeded.
Typical
5
55
-
3
60
5
60
458
Max
Units
20
µA
60
mA
-32
Vdc
3.6
Vdc
65
mA (1)
18
Vdc
50
mV ac pk-pk
65
mA (1)
Battery Life
14 days
1 year
T24 Telemetry User Manual

Advertisement

Table of Contents
loading

This manual is also suitable for:

T24-hs-lsT24-acmi