Profimess SW-08 Operating Instructions Manual page 202

Varible area flowswitch
Table of Contents

Advertisement

Operation
Switch point adjustment
Setting the switch point of a non-
installed device
Fig. 32: Loosen the set screws
1.
Loosen the set screws of the
switch contact (Fig. 32/1) using
a flat-bladed screwdriver.
Set the switch contact to the flow
2.
value to be monitored. Make
sure that the arrow on the switch
contact data plate is exactly in
alignment with the desired flow
rate on the switchpoint adjust-
ment scale.
Re-tighten the set screws of the
3.
switch contact (Fig. 32/1) using
a flat-bladed screwdriver. When
doing so, observe the correct
tightening torque of the screws.
Ä
Chapter 11.1 "Tightening
torque" on page 79
44
ð The set switch point corre-
sponds to the switch-off
point of the switch contact
by decreasing flow.
20.02.2015

Advertisement

Table of Contents
loading

Table of Contents