positioning aid label and the
label on the device body must
be removed and repositioned.
Parallax errors are to be
avoided.
20.
Loosen the switch contact set
screws and adjust the switch
contact to a graduation mark in
the middle of the scale on the
device body. Now tighten the set
screws, observing the proper
torque (
Ä
Chapter 11.1 "Tight-
ening torque" on page 78).
Hold the device vertically and
21.
with the test rod, move the float
in the flow direction until the
switch contact switches (con-
tinuity meter). Now slowly return
the float to its original position
and, with the aid of the continuity
meter, determine at which scale
value the switch contact
switches off.
22.
Adjust the display base plate
with attached plastic housing
and scale so that the cut-off
value on the display scale corre-
sponds to the value on the scale
on the device body. Align the
display base plate perpendicular
to the device body.
If the value on the display scale
23.
matches the value on the scale
on the device body, secure the
mounting screws of the display
base plate, observing the proper
torque (
Ä
Chapter 11.1 "Tight-
ening torque" on page 78).
23.02.2015
Fig. 60
Fit the display cover plate onto
24.
the plastic housing and carefully
press down (Fig. 60) until it
snaps into place.
Fig. 61
25.
Screw in and tighten the fixing
screw of the display housing
cover (Fig. 61), observing the
proper torque (
"Tightening torque" on page 78).