Test Channel Output - Teledyne 400E Instruction Manual

Ozone analyzer
Table of Contents

Advertisement

Model 400E Ozone Analyzer
Instruction Manual

6.5.7. TEST Channel Output

When activated the A4 Analog Output channel can be used to report the real-time
value of one of several of the Test Function viewable from the SAMPLE mode
display. To activate the A4 channel and select a Test Function press:
P/N 04316 Rev: B
SAMPLE
< TST TST > CAL
Press Exit at
Any Time to
SETUP X.X
Return to Main
SAMPLE
CFG DAS RNGE PASS CLK
Display
SETUP X.X
Press Exit at
Any Time to
COMM VARS
Return to
SETUP Menu
DIAG
DIAG
PREV
Press Exit at
Any Time to
Return to DIAG
DIAG
Menu
PREV
DIAG TCHN
DIAG TCHN
MEASURE
PREV
Press PREV or
NEXT to move up
and down list of
functions
(See Table 6-7)
RANGE = 500.0 PPB
MORE
DIAG
HALT
SIGNAL I / O
NEXT
ENTR
ANALOG OUTPUT
NEXT
ENTR
Continue pressin NEXT until ...
TEST CHAN OUTPUT
NEXT
ENTR
TEST CHANNEL: NONE
NEXT
ENTR
TEST CHANNEL: PHOTO
NEXT
ENTR
Press ENTR to
select the Function
on the display and
activate the Test
Channel
Operating Instructions
O3 =XXX.X
SETUP
EXIT
EXIT
EXIT
EXIT
EXIT
EXIT
EXIT
Press EXIT to
return to the
DIAG Menu
79