Figure 3-4: Pneumatic Connections Flow Diagram With Zero/Span Valve Or Izs Options Installed - Teledyne 400E Instruction Manual

Ozone analyzer
Table of Contents

Advertisement

Getting Started
Sample Gas pressure must equal ambient atmospheric
In applications where the sample gas is received from a
pressurized manifold, a vent must be placed as shown to
equalize the sample gas with ambient atmospheric pressure
If your analyzer is equiped with wither the Zero/Span Valve Option (Option 50) or
the Internal Zero/Span Option (IZS - Option 51), the pneumatic connections should
be made as follows:
Figure 3-4: Pneumatic Connections Flow Diagram with Zero/Span Valve or
Once the appropriate pneumatic connections have been made, check all pneumatic
fittings for leaks using the procedures defined in Section 9.3.4.
32
before it enters the analyzer.
This vent line must be:
At least 0.2m long
q
No more than 2m long
q
Vented outside the shelter or immediate
q
area surrounding the instrument
MODEL 700
Gas Dilution Calibrator
(w/ Photometer Option)
or
VENT
MODEL 401
Ozone Generator
MODEL 701
Zero Air Generator
MODEL 701
Zero Air Generator
IZS Options Installed
NOTE
pressure.
Option 50 – Zero/Span Valves
Source of
SAMPLE Gas
VENT if input is pressurized
Sample
Exhaust
Span
Zero Air
VENT
Dry Air
Option 51 - Internal Zero/Span Option
Source of
VENT if input is pressurized
SAMPLE Gas
Sample
Exhaust
Span
Zero Air
Dry Air
VENT
Model 400E Ozone Analyzer
Instruction Manual
MODEL
400E
(IZS)
MODEL
400E
P/N 04316 Rev: B