Property
NAME
The data channel's name (for reports and RS-
232 access).
TRIGGERING
The event that triggers this data channel to
EVENT
measure and store its data parameters.
See APPENDIX A for a list of available
triggering events.
LIST OF
A User-configurable list of data types to be
PARAMETERS
recorded.
See APPENDIX A for a list of available
parameters.
REPORT PERIOD
The amount of time between each report.
NUMBER OF
The number of reports that will be stored in the
RECORDS
data file.
RS-232 REPORT
Indicates whether or not a report will be
printed automatically to the RS-232 channel.
CHANNEL
Provides a convenient means to temporarily
ENABLED
disable a data channel.
CAL. HOLD OFF
Disables sampling of data parameters while
instrument is in calibration mode.
6.10.3.2. Data Parameters
Data Parameters are actual types of the data may be measured and stored by the
iDAS. For each of T-API's instrument models, the list of available data parameters is
set and non-customizable (see Appendix A for a list of available Data Parameters
for this instrument).
Most data parameters have measurement units associated with them, such as mV,
PPB, cc/m, etc., although some have no units. The iDAS is not designed to permit
these units to be changed.
Each Data Parameters has user-configurable functions that define how the data is
recorded.
106
Table 6–20: iDAS Data Channel Properties
Description
Model 300E Carbon Monoxide Analyzer
Initial
Setting
"NONE"
ATIMER
1
000:01:00
100
OFF
ON
OFF
Instruction Manual
Setting Range
Up to 6 letters and
digits.
Any single event
chosen form the list
Only limited by
memory space
000:00:01 to
366:23:59
(Days:Hours:Min.)
1 to 999,999
(limited by available
storage space)
OFF or ON
OFF or ON
OFF or ON
P/N 04288 REV: A