Required facility needs

NOTE:
2
1
1. Door – at least 762 mm (30
inches)
2. Film Viewer
3. Counter Top, Sink with hot
and cold water and Supplies
Storage
4. Linen Supply
5. Probes/Supplies
Figure 3-1. Minimal floor plan, 2.5 m x 3 m (8 by 10 foot)
Ultrasound System
Direction 5444964-100 English Rev. 5
GE requires a dedicated power and ground for the proper
operation of its Ultrasound equipment. This dedicated power
shall originate at the last distribution panel before the Ultrasound
system.
The Ultrasound system will function on voltages from 100-240
Volts and 50 or 60 Hz. However, if using 220-volt power in North
America, then a center tapped power source is required.
Sites with a mains power system with defined Neutral and Live:
The dedicated line shall consist of one phase, a neutral (not
shared with any other circuit), and a full-size ground wire from
the distribution panel to the Ultrasound outlet.
Sites with a mains power system without a defined Neutral:
The dedicated line shall consist of one phase (two lines), not
shared with any other circuit, and a full-size ground wire from
the distribution panel to the Ultrasound outlet.
3
9
8
7
6
6. Examination Table – 1930 x
610 mm (76 x 24 inches)
7. Footswitch
8. Stool
9. Ultrasound system
10. External Peripherals
11. Dedicated Power Outlet -
Circuit Breaker protected and
easily accessible
Common Service Information
4
5
11
10
12
13
14
Facility needs
Scale:
Each square equals one square
foot (app. 31 x 31 cm)
12. Network Interface
13. 457 mm (18 inches) distance
of Ultrasound system from
wall or objects
14. GE Cabinet for Software and
Manuals
3-23