Bep2 - Ecg Module Installation Procedure - GE Vivid 4 BT03 RFI Service Manual

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
2392751-100, R
IRECTION
8-5-13-4

BEP2 - ECG Module Installation Procedure

NOTE:
BEP2 refers to a Pentium 4-based BEP (newer type).
1.) Slide the new ECG Module into the front of the BEP, pushing it in until it aligns with other modules
at the front of BEP.
2) Secure the ECG Module with four screws, two from each side of the BEP (see
3.) Replace the BEP Front Panel and secure it with the two screws previously removed.
4) Reconnect the power cable and the COM 2 cable to the rear of the new ECG Module.
5) Install the BEP left and right side covers, as described in the
Procedure
6.) Carefully return the BEP to its original location and secure all brackets by tightening the screws.
7) Reconnect all cables at the rear of the BEP.
8) Return the Vivid™ 4 rear cover to its original position - see
on page 8 - 7.
9) Return the Vivid™ 4 side covers, as described in the
10) Reconnect the mains AC power cable from the wall outlet.
11) Turn ON the main circuit breaker located on rear of the Vivid™ 4.
12) Turn ON the Vivid™ 4 by pressing the ON/OFF button on the Control Console.
8-72
3
EVISION
Figure 8-55 ECG Module inside BEP2
on page 8 - 54.
Section 8-5 - Back End Parts Replacement
V
IVID
BEP2 - Side Cover Installation
Rear Cover Replacement Procedures
Side Covers Installation Procedure
™ 4 S
M
ERVICE
ANUAL
Figure
8-54, above).
on page 8 - 4.