Upper Rear Cover Replacement Procedure - GE LOGIQ 9 Technical Manual

Hide thumbs Also See for LOGIQ 9:
Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
2294854-100, R
IRECTION
8-2-3

Upper Rear Cover Replacement Procedure

8-2-3-1
Manpower
One Person, 10 minutes
8-2-3-2
Tools
Phillips screwdriver
8-2-3-3
Preparations
Power Down/Shutdown the system as described in
8-2-3-4
Filter Cover Removal Procedure
Take hold of the bottom side of the Filter Cover. Push up, pull out and unhook it from the top side and
remove it.
8-2-3-5
Filter Cover Installation Procedure
(click here to view the cover installation video)
Align the cover at the top side. Push in and down to secure it along the bottom.
PROPRIETARY TO GE
3
EVISION
Select the movie camera icon to view the video of the Cover
removal procedure.
Installation starts at the upper side
Figure 8-3 Filter Cover
Chapter 8 Replacement Procedures
LOGIQ™ 9 P
Section 4-3-2 on page
4-3.
Filter Cover
Removal starts from the lower side
M
ROPRIETARY
ANUAL
8-7