Warning:
toxic materials, and to avoid spreading contamination to
uncontaminated surfaces, wear clean, chemical-resistant,
powder-free gloves when working with the source components.
To remove the probe adjuster assembly probe and source enclosure seals:
1.
Remove the source enclosure from the instrument (see
2.
Use the O-ring removal kit to carefully remove the following seals from
the probe adjuster assembly (see
•
Probe seal
•
Nebulizer gas seal
3.
Use the O-ring removal kit to carefully remove the following seals from
the source enclosure:
•
Source enclosure seal
•
Nebulizer gas seal
•
Desolvation gas seal
To avoid personal contamination with biohazards or
page
Probe adjuster nebulizer gas seal
Probe adjuster assembly probe seal
March 14, 2014, 715004387 Rev. B
Replacing the source assembly seals
page
99):
90).
175