Waters Xevo TQD Overview And Maintenance Manual page 175

Hide thumbs Also See for Xevo TQD:
Table of Contents

Advertisement

Warning:
toxic materials, and to avoid spreading contamination to
uncontaminated surfaces, wear clean, chemical-resistant,
powder-free gloves when working with the source components.
To remove the probe adjuster assembly probe and source enclosure seals:
1.
Remove the source enclosure from the instrument (see
2.
Use the O-ring removal kit to carefully remove the following seals from
the probe adjuster assembly (see
Probe seal
Nebulizer gas seal
3.
Use the O-ring removal kit to carefully remove the following seals from
the source enclosure:
Source enclosure seal
Nebulizer gas seal
Desolvation gas seal
To avoid personal contamination with biohazards or
page
Probe adjuster nebulizer gas seal
Probe adjuster assembly probe seal
March 14, 2014, 715004387 Rev. B
Replacing the source assembly seals
page
99):