Waters Xevo TQD Overview And Maintenance Manual page 157

Hide thumbs Also See for Xevo TQD:
Table of Contents

Advertisement

4.
Use the needle-nose pliers to slide a new liner sleeve and ferrule onto
the capillary.
5.
Insert the capillary in the PEEK union, and ensure that it is fully
seated.
6.
Screw the UNF coupling into the PEEK union, finger-tight only.
7.
Gently tug on the capillary, testing to ensure that it stays in place.
8.
Use the 7-mm wrench to tighten the locknut against the PEEK union
until the union can no longer be twisted.
9.
Slide a new conductive sleeve and the knurled collar over the capillary.
10. Tighten the knurled collar onto the UNF coupling.
Warning:
wear safety goggles when performing the leak test.
11. Perform a leak test by attaching the free end of the PEEK tubing to an
LC pump and pumping 50:50 acetonitrile/water through it at 1 mL/min.
If leakage occurs, disassemble and remake the connection, and then
repeat the leak test.
If the backpressure on the LC pump is high, replace the capillary,
and repeat the leak test.
12. When no leakage occurs and the backpressure on the LC pump is
normal, disconnect the PEEK tubing from the LC pump.
13. Remove the probe inlet connector and PEEK tubing from the PEEK
union.
14. Carefully thread the capillary through the probe assembly.
To avoid eye injury from high-pressure liquid jet spray,
March 14, 2014, 715004387 Rev. B
Replacing the ESI probe sample capillary
157

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents