Warning:
toxic materials, and to avoid spreading contamination to
uncontaminated surfaces, wear clean, chemical-resistant,
powder-free gloves when working with the probe and source
components.
To install the new capillary:
1.
Use the sharp knife or PEEK tubing cutter to cut a piece of red PEEK
tubing approximately 60 cm (24 inches) long.
Requirement:
horizontal axis).
2.
Insert one end of the red PEEK tubing in the probe inlet connector, and
screw the connector, finger-tight, into the PEEK union.
Doing so ensures a minimum dead volume when fitting the
Rationale:
capillary.
3.
Fit the UNF coupling to the new capillary.
Requirement: Use a UNF coupling with no grooves, which is appropriate
for the IonSABRE II probe.
Replacing the IonSABRE II probe sample capillary
To avoid personal contamination with biohazards or
Cut the tubing squarely (that is, perpendicular to its
Probe inlet connector
PEEK tubing
TP02671
March 14, 2014, 715004387 Rev. B
163