4 Maintenance Procedures
Warning:
toxic materials, and to avoid spreading contamination to
uncontaminated surfaces, wear clean, chemical-resistant,
powder-free gloves when working with the probe and source
components.
Warning:
the ESI probe tip is sharp.
To install the new capillary
1.
Use the sharp knife or PEEK tubing cutter to cut a piece of red PEEK
tubing approximately 60 cm (24 inches) long.
Requirement:
horizontal axis).
2.
Insert one end of the red PEEK tubing in the probe inlet connector, and
screw the connector into the PEEK union so that it is finger-tight.
Rationale:
capillary.
3.
Fit the UNF coupling to the new capillary.
156
March 14, 2014, 715004387 Rev. B
To avoid personal contamination with biohazards or
To avoid puncture wounds, handle the probe with care;
:
Cut the tubing squarely (that is, perpendicular to its
Doing so ensures a minimum dead volume when fitting the
Probe inlet connector
PEEK tubing
TP02671