Waters Xevo TQD Overview And Maintenance Manual page 75

Hide thumbs Also See for Xevo TQD:
Table of Contents

Advertisement

To prevent the sprayer from colliding with the cone and
Notice:
breaking, always retract the stage before installing the source
enclosure or closing the door.
4.
On the NanoFlow source, loosen the stage retaining screw, pull the stop
screw, and slide the stage fully out of the enclosure.
Stop screw
5.
Using both hands, fit the NanoFlow source enclosure to the two
supporting studs on the source adaptor housing.
6.
Close the source enclosure door.
7.
Connect a 1/16-inch PTFE tube between the mass-flow controller output
(mounted beneath the stage on the front of the NanoFlow source) and
your sprayer.
For details regarding how to fit each sprayer, see the corresponding
Tip:
reference:
Waters Universal NanoFlow Sprayer Installation and Maintenance
Guide (part number 71500110107)
"Fitting a borosilicate glass capillary (nanovial)" on
Capillary Electrophoresis/Capillary Electrochromatography
Sprayer User's Guide (part number 6666522)
Retaining screw
March 14, 2014, 715004387 Rev. B
NanoFlow source
page 77
75

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents