3 Changing the Mode of Operation
Warning:
or stainless steel "finger tight" screws to connect the selector valve
to the ESI probe; use the PEEK tubing and natural (beige) colored
PEEK "finger tight" screws supplied with the instrument.
6.
Using PEEK tubing equal to 0.004-inch ID, connect port S of the selector
valve to the ESI probe.
Recommendation:
for sample flow rates ≤1.2 mL/min; use 0.005-inch ID tubing for sample
flow rates >1.2 mL/min.
Requirements:
If you are replacing the tubing between the selector valve and the
probe, minimize the length to reduce peak broadening.
When cutting the tubing to length, cut it squarely (that is,
perpendicular to its horizontal axis).
60
March 14, 2014, 715004387 Rev. B
To avoid electric shock, do not use stainless steel tubing
To reduce peak broadening, use 0.004-inch ID tubing