Requirement:
squarely (that is, perpendicularly to its horizontal axis).
8.
Route the open end of the exhaust tubing to a suitable exhaust vent.
9.
Connect electric cables to the roughing pump (see
Connecting electric cables to the Edwards oil-free roughing pump
Roughing pump connections:
To connect cables:
1.
Connect the relay cable from the roughing pump's d.c. connector to the
pump connector on the mass spectrometer's rear panel.
2.
Connect the roughing pump power cord to the main power source.
Connecting the Edwards oil-free roughing pump
Use the sharp knife to cut the PVC exhaust tubing
Roughing pump
connector
To power source
page
Roughing pump
d.c. connector
March 14, 2014, 715004387 Rev. B
227).
TP03165