4 Maintenance Procedures
18. Use the combined 2.5-mm Allen wrench and cone extraction tool to
remove the captive screws securing the 2 PEEK extraction cone retainer
clips.
Notice:
• To avoid damaging the extraction cone aperture, take great
• To avoid damaging the fragile extraction cone, do not place it
19. Remove the extraction cone from the ion block.
132
March 14, 2014, 715004387 Rev. B
care when removing the extraction cone from the ion block.
on its tip; instead, place it on its flanged base.
Extraction cone aperture
Securing screw
Retainer clip
Extraction cone