B External Connections

Connecting to the collision cell gas supply

Required materials
Chemical-resistant, powder-free gloves
Wrench
1/8-inch Swagelok
1/8-inch stainless steel tube (supplied with the instrument)
To connect the collision cell gas supply:
1.
Use the 1/8-inch Swagelok nut and ferrule to connect the 1/8-inch
stainless steel tube to the collision cell gas inlet on the rear of the
instrument (see the figure on
2.
Use the wrench to tighten the 1/8-inch Swagelok nut.
3.
Connect the free end of the tube to the collision gas supply.

Connecting the nitrogen exhaust line

Required materials
Chemical-resistant, powder-free gloves
Sharp knife
10-mm and 12-mm PTFE tubing (included in the Waters Rough Pump
Connect Kit)
®
snoop
230
March 14, 2014, 715004387 Rev. B
®
nut and ferrule
(or equivalent) leak detector liquid
page
229).