Advertisement

4 Maintenance Procedures
Maintenance schedule:
Procedure
Clean the IonSABRE II probe
tip. (Options using the
IonSABRE II probe only.)
Replace the IonSABRE II
probe capillary. (Options
using the IonSABRE II probe
only.)
Clean or replace the corona
pin (APCI and ESCi modes).
Replace the IonSABRE II
probe heater. (Options using
the IonSABRE II probe only.)
Replace the ion block heater
cartridge.
Replace the source assembly
seals.
Replace the instrument's air
filters.
Change the roughing pump
oil.
84
March 14, 2014, 715004387 Rev. B
Frequency
When sensitivity
decreases to
unacceptable levels or
when significant
chemical interference
is present.
When sensitivity
decreases to
unacceptable levels or
sample flow is
inconsistent.
When the corona pin is
corroded or black, or
the sensitivity
decreases to
unacceptable levels.
If the heater fails to
heat when the
instrument is pumped
down (evacuated).
If the heater fails to
heat when the
instrument is pumped
down (evacuated).
Annually.
Annually.
Annually.
For information...
See
page
159.
See
page 160.
See
page
166.
See
page 167
See
page
170.
See
page
174.
See
page
178.
See
page
180.