Waters Xevo TQD Overview And Maintenance Manual page 161

Hide thumbs Also See for Xevo TQD:
Table of Contents

Advertisement

4.
Remove the end cover and gasket.
Nebulizer adjuster knob
5.
Unscrew and remove the nebulizer adjuster knob.
6.
Remove the PEEK union/UNF coupling assembly and capillary from the
probe.
The PEEK union used with the IonSABRE II probe is notched on
Tip:
one of its flats, a feature that distinguishes it from the PEEK union used
with the ESI probe (see "Replacing the ESI probe sample capillary" on
page
151).
PEEK union/UNF coupling assembly
Notch
Locknut
Replacing the IonSABRE II probe sample capillary
Capillary
March 14, 2014, 715004387 Rev. B
Gasket
End-cover
161