Waters Xevo TQD Overview And Maintenance Manual page 220

Hide thumbs Also See for Xevo TQD:
Table of Contents

Advertisement

B External Connections
Bear in mind these requirements when connecting the roughing pump:
The pump must be horizontal, or within one degree of horizontal.
The ambient temperature of the area in which you place the pump must
range between 15 and 40 ºC.
The pump must be oriented so that it permits easy access to the gas
ballast valve and oil-level sight glass.
For proper ventilation, the following minimum clearances must apply:
Left-side minimum clearance
is 15.24 cm (6 inches)
Front-side minimum clearance
is 35.56 cm (14 inches)
To connect the roughing pump:
1.
Place the PTFE drip tray on the floor, within 5 feet of the instrument.
Warning:
must lift the roughing pump.
2.
Place the pump on the PTFE drip tray.
220
March 14, 2014, 715004387 Rev. B
To avoid personal contamination with biohazards
Warning:
or toxic materials, and to avoid spreading contamination to
uncontaminated surfaces, wear clean, chemical-resistant,
powder-free gloves when working with the pump and its
connections.
To avoid skeletal or muscle injury, at least two people
Back-side minimum clearance
is 15.24 cm (6 inches)
TP02689
Right-side minimum clearance
is 15.24 cm (6 inches)